Detailed description:
Product Name: medical infrared thermometer
Product model: CH-818
Measurement mode: frontal temperature and ear temperature
Measuring distance frontal temperature: 1cm ~ 3cm
Temperature measurement range: 32.0 * C ~ 43.0 * C
Maximum allowable error: the thermometer is within the temperature display range of 32,0 ° C ~ 34.9 * C, and the maximum allowable error is 0,3 ° C
The thermometer is within the temperature display range of 35.0 ° C ~ 42.0 * C, and the maximum allowable error is ± 0.2 * C
The thermometer is within the temperature display range of 42.1 * C ~ 43.0 * C, and the maximum allowable error is 0.3 * C
Temperature resolution 0.1 * C / 0.1 "F
Operating ambient temperature: 10 * C-40 ° C relative humidity: ≤ 80% atmospheric pressure: 860hpa ~ 1060hpa
Transportation and storage conditions temperature: 20.0 * C - + 55 * C relative humidity: ≤ 93% atmospheric pressure: 500hPa ~ - 1060hpa
Power supply DC3.0V (2 AAA batteries)
Automatic shutdown time: 35 seconds
Power prompt 2.4V earth 0.2V (prompt voltage at low power)
Size 155mmx38mmx36mm
Weight ≤ 115g (excluding battery)
When the operating range of the ambient temperature of the prompt function exceeds 10 ° C ~ 40 ° C, the thermometer shall have a prompt signal or stop providing the temperature value.
In body temperature mode, Lo "C" is displayed if the measured temperature is less than 32.0c; If it is greater than 43.0 * C, hi "C prompt signal is displayed.
One key measurement, it is simple and convenient to gently press the measurement key. Parents and babies can operate easily. Non-contact forehead and ear temperature measurement is dual-mode, and the measurement distance is 1-3cm. It does not touch human skin to prevent cross infection.